Low Fio; Alarm; High Spo - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

6.3.20

Low FiO

Property
Alarm text
Priority
Alarm condition
Possible cause
Setting range
Setting resolution
Ventilator action
6.3.21

High SpO

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
130
Alarms
User manual

Alarm

2
Description
Low FiO2
High
A Low FiO
below the alarm limit for 30 seconds.
Decreased oxygen inlet.
Disconnection of oxygen inlet.
Increased minute ventilation.
High leakage.
21% to 100%
Off
1%
The ventilator will continue treatment with the same settings.
Alarm
2
Description
High SpO2
Medium
A High SpO
exceeds the alarm limit for 30 seconds.
Too high flow of bleed-in oxygen.
The SpO
The Vivo 45 LS will continue treatment according to the
current settings.
90 % to 100 %
Off
1 %
alarm will be given when the measured FiO
2
alarm will be given when the measured SpO
2
value goes back below the alarm limit.
2
is
2
2
Doc. 007301 K-1