6.3.11
Low MV
Property
Alarm text
Priority
Alarm condition
Possible cause
Setting range
Setting resolution
Ventilator action
100
Alarms
User manual
(Low Inspired Minute Volume Alarm)
i
Description
Low MVi
High
A Low Inspired Minute Volume alarm will be given when the
monitored minute volume does not reach the alarm limit for
15 seconds.
This alarm is only used if the Vivo 65's patient circuit type is set
to Exhalation valve (single limb) circuit.
•
Mismatch between Breath Rate and Inspired Tidal Volume
settings and the alarm setting.
•
Changes in airway resistance and or compliance.
•
Decreased Breath Rate.
•
1.0 l/min to 30 l/min (Adult mode)
•
0.1 l/min to 10 l/min (Paediatric mode)
•
Off
0.1 l up to 1.0 l, 0.5 l above 1.0 l.
The ventilator will continue treatment with the same settings.
Doc. 006150 M-1