6.3.14
Low Breath Rate Alarm
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
Doc. 007301 K-1
Description
Low Breath Rate
High
A Low Breath Rate alarm will be given when the delivered
total breath rate is below the alarm limit for 15 seconds.
•
Mismatch between the Breath Rate setting and the alarm
setting.
•
The patient cannot trigger breaths because the inspiratory
trigger setting is too high.
•
Decrease in the patient's spontaneous breathing.
•
Circuit disconnection.
The breath rate goes above the alarm limit.
The Vivo 45 LS will continue treatment according to the
current settings.
•
4 bpm to 30 bpm (Adult mode, non-MPV modes)
•
1 bpm to 30 bpm (Adult mode MPV modes)
•
6 bpm to 50 bpm (Pediatric mode)
•
Off
1 bpm.
Alarms
125
User manual