Low Breath Rate Alarm

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
Doc. 007301 K-1
Description
Low Breath Rate
High
A Low Breath Rate alarm will be given when the delivered
total breath rate is below the alarm limit for 15 seconds.
Mismatch between the Breath Rate setting and the alarm
setting.
The patient cannot trigger breaths because the inspiratory
trigger setting is too high.
Decrease in the patient's spontaneous breathing.
Circuit disconnection.
The breath rate goes above the alarm limit.
The Vivo 45 LS will continue treatment according to the
current settings.
4 bpm to 30 bpm (Adult mode, non-MPV modes)
1 bpm to 30 bpm (Adult mode MPV modes)
6 bpm to 50 bpm (Pediatric mode)
Off
1 bpm.
Alarms
125