6.3.10
Low Vt
Property
Alarm text
Priority
Alarm Condition
Possible cause
Reset criteria
Setting range
Setting resolution
Ventilator action
Doc. 007301 K-1
Alarm (Low Expired Tidal Volume)
e
Description
Low Vte
High
A Low Expired Tidal Volume alarm will be given when the
monitored Expired Tidal Volume fails to reach the set limit
for the Low Expired Tidal Volume alarm for 15 seconds.
Mismatch between Expired Tidal Volume and Alarm setting.
Changes in airway resistance and or compliance.
Leakage around the mask or within one of the components
of the circuit.
Full breath above set alarm limit
100 ml to 2000 ml (Adult mode)
20 ml to 500 ml (Pediatric mode)
Off
10 below 600 ml, 100 above 600 ml
The ventilator will continue treatment with the same settings.
Alarms
121
User manual