Ambient pressure range:
700 to 1100 mbar, corresponding to
~4200 meters (13800 feet) above sea
level to ~700 meters(2300 feet) below
sea level, at normal atmospheric
pressure.
As seen in the graph above, the ventila-
tor is unable to deliver set max pressure
at a very low ambient pressure.
Ingress Protection:
IP22
Solid particle protection: Hazardous parts are protected from touch by fingers and by
objects greater than 12 mm.
Liquid ingress protection: The protection withstands dripping water less than 15 degrees
from vertical.
The ingress protection has been tested by water drips equivalent to 3mm rain/minute
for 10 minutes (2.5 minutes for each tilting direction).
8.2.6

Other

Patient Circuit Leakage
Recommended leakage: 20 to 50 l/min at 10 cmH
Minimum leakage: 12 l/min at 4 cmH
Oxygen Inlet
Oxygen inlet port: Maximum flow: 15 l/min (medical oxygen). Oxygen coupling is
type CPC PMCD181032.
Start-up Time
Start-up from unpowered state: about 20 seconds.
Sound Power Level
Sound level at 10 cmH
Alarm sound level : Adjustable 50–80 dB(A), Measured at 1m. Tolerance: ± 5 dB(A).
Doc. 007301 K-1
O (leakage circuit)
2
O in CPAP mode: Less than 30 dB(A). Measured at 1 m.
2
O (leakage circuit)
2
Technical Specifications