6.3.2
Low Pressure Alarm
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
Setting display
Doc. 007301 K-1
Description
Low Pressure
High
A Low Pressure alarm will be given when the Vivo 45 LS
pressure fails to reach the low pressure alarm limit for 15
seconds.
•
Disconnection of patient circuit.
•
Mismatch between pressure setting and alarm setting.
•
Leakage from the mask or other components of the patient
circuit.
The pressure rises above the alarm limit.
The Vivo 45 LS will continue treatment according to the
current settings.
•
1 cmH
O to 50 cmH
2
Note that the Low pressure alarm cannot be set higher than
the value set for the High pressure alarm.
Below 10 cmH
O: 0.5 cmH
2
Above 10 cmH
O: 1.0 cmH
2
The Low Pressure alarm setting is displayed by a red line in
the pressure bar graph.
O
2
O
2
O
2
Alarms
113
User manual