Low Pressure Alarm - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

6.3.2

Low Pressure Alarm

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
Setting display
Doc. 007301 K-1
Description
Low Pressure
High
A Low Pressure alarm will be given when the Vivo 45 LS
pressure fails to reach the low pressure alarm limit for 15
seconds.
Disconnection of patient circuit.
Mismatch between pressure setting and alarm setting.
Leakage from the mask or other components of the patient
circuit.
The pressure rises above the alarm limit.
The Vivo 45 LS will continue treatment according to the
current settings.
1 cmH
O to 50 cmH
2
Note that the Low pressure alarm cannot be set higher than
the value set for the High pressure alarm.
Below 10 cmH
O: 0.5 cmH
2
Above 10 cmH
O: 1.0 cmH
2
The Low Pressure alarm setting is displayed by a red line in
the pressure bar graph.
O
2
O
2
O
2
Alarms
113
User manual