Breas Vivo 45 LS User Manual page 166

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

Rise Time
Range/Performance:
1 to 9 (PSV, PCV)
Resolution:
1
Inspiratory Trigger
Range/Performance: 1 to 9 (PSV), 1 to 9, Off (PCV).
Resolution: 1
SIMV Support Pressure
Range/Performance: 4 to 50 cmH
Tolerance: ±0.5 cmH
Resolution: 0.5 below 10 cmH
Expiratory Trigger
Range/Performance: 1 to 9.
Resolution: 1
Min Inspiration Time
Range/Performance:
Off, 0.3 to 3 s (Adult), Off, 0.3 to 2 s (Pediatric)
Resolution: 0.1 s
Max Inspiration Time
Range/Performance:
0.3 to 5 s, Off (Adult), 0.3 to 2 s, Off (Pediatric)
Resolution: 0.1 s
Backup Rate
Range/Performance:
4 to 30 bpm (Adult), 6 to 60 bpm (Pediatric). Tolerance: ±2%
Resolution: 1 bpm
166
Technical Specifications
User manual
O.
2
O or ±5%, whichever is greatest.
2
O, 1.0 above 10 cmH
2
O
2
Doc. 007301 K-1