Charging The Batteries - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

2 Check that the:
• Click-in battery power source LED is illuminated
• Medium priority "Lost Mains Power" alarm is triggered
• Information message "Switched to Click-In Battery" is posted
External DC Switchover Test
Test Conditions
1 Disconnect the AC power cord.
2 Check that the:
• External DC power source LED is illuminated
• Medium priority "Lost Mains Power" alarm is triggered
• Information message "Switched to External DC" is posted
5.6.3

Charging the Batteries

CAUTION!
Do not charge the Vivo 45 LS while placed in the carry bag or other types of closed
or non-ventilated spaces.
The internal and click-in batteries are automatically charged when connecting the venti-
lator to the mains supply. To ensure that the batteries are fully charged, a maintaining
charging cycle will be performed. The batteries are not charged when connecting the
ventilator to an external DC supply. While charging, the battery level will be animated.
The batteries are only charged if the internal temperatures are between 0 to 45°C (32 to
113°F). High power consuming settings in combination with high ambient temperatures
may make the battery temperature rise above 45°C (113°F).
NOTE
Charging of batteries is only started when the state of charge is below 95%
Behaviour of the ventilator while Internal Battery or Click-in Battery is
Charging
The ventilator can operate as normal while a battery is charging. The battery icon will be
animated (filling in from bottom to top) while charging.
Doc. 007301 K-1
AC power connected
External DC connected
Treatment is running
How to Use the Vivo 45 LS
User manual
91