Internal/Click-In Battery Hot Alarm; Internal Function Failure - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

6.4.22

Internal/Click-In Battery Hot Alarm

Property
Alarm text
Alarm condition
Priority
Ventilator action
6.4.23

Internal Function Failure

Property
Alarm text
Priority
Alarm condition
Reset criteria
Ventilator action
Action to take
Doc. 007301 K-1
Description
Internal Battery — Internal Battery Hot
Click-In Battery — Click-In Battery Hot
An alarm for Internal/Click-In Battery Overheat in Dis-
charge will be given when the internal or click-in battery rea-
ches 55°C (131°F).
NOTE
The battery electronics by manufacture stops
discharge at 60°C (140°F).
High
The ventilator will continue treatment with the same settings.
Battery discharging will be disabled (by the battery elec-
tronics) once the temperature gets to 60°C (140°F). (If the
battery is last power source, the ventilator will stop running).
Description
Int. Function Failure
High
Failure of internal function that prevents treatment or nor-
mal operation of the ventilator.
The error code that follows the alarm text indicates the kind
of function failure.
All Internal Function Failure alarm error codes are defined
and explained in the ventilator Service Manual.
Restart the ventilator.
The ventilator will stop the treatment and shut down.
Restart the Vivo 45 LS.
If the alarm persists or reoccurs: Take a note of the error
code and contact your supplier of the Vivo 45 LS .
Alarms
147
User manual