5.5.21
Min Pressure
Item
Definition
Setting min
Setting max
Setting
resolution
Tolerance
Default
value
5.5.22
CPAP
Item
Definition
Setting min
Setting max
Setting
resolution
Max bias
error
Default
value
5.6
Using Batteries
Since all batteries, in general, degenerate over time, the recommendations below will
ensure that the battery capacity of the ventilator is maximized during its lifetime.
The internal and click-in batteries in the ventilator are of the Lithium-ion type, which is
a high performance battery. It has long expected lifetime, low weight in relation to its
Doc. 007301 K-1
Description
The Min Pressure setting is only used when Target Volume is acti-
vated. Min Pressure defines the lower pressure limit down to where
the ventilator can decrease the pressure to maintain the set Target Vol-
ume. If the actual volume is above Target Volume at Min Pressure,
the ventilator will continue to ventilate at this Min Pressure setting.
4 cmH
O
2
Current Max pressure setting
0.5 below 10 cmH
O, 1.0 above 10 cmH
2
±0.5 cmH
O or ±5 %, whichever is the greatest.
2
15 cmH
O
2
Description
The CPAP setting defines the pressure that will be applied to the air-
ways in CPAP mode.
4 cmH
O
2
20 cmH
O
2
0.5 below 10 cmH
O,
2
1.0 above 10 cmH
O
2
±0.8 cmH
O or ±4 %,whichever is the greatest.
2
10 cmH
O (Adult mode)
2
8 cmH
O (Pediatric mode)
2
O
2
How to Use the Vivo 45 LS
User manual
89