5.3.12

Spont Rate

The Spont Rate displays the actual spontaneous breath rate.
5.3.13

% Spont

% Spont displays the percentage of spontaneous breaths calculated since the ventilator
was last started (not calculated until 100 breaths are registered).
5.3.14
SpO
The SpO
accessory.
1. Peak Wavelength (red): 660 nm
2. Peak Wavelength (infrared): 905 nm
3. Maximum Optical Output Power: 15 mW
4. For more information regarding the oxygen probe's range of peak wavelengths, max
optical power and usage, please refer to the respective probe manual.
5.3.15

Pulse Rate

Pulse Rate displays the patient's pulse rate as measured with the SpO
5.3.16

I:E

I:E displays the ratio between the length of the inspiration and the length of the
expiration.
5.3.17

Insp. Time

Insp. Time displays the duration of the inspiratory cycle, measured from the start of
inspiration to the start of expiration.
Doc. 007301 K-1
(Oxygen Saturation)
2
displays the patient's oxygen saturation, if measured with the SpO
2
When using the ventilator with the SpO
functional oxygen saturation measured by the sensor.
The following information concerns the light emitted by the SpO2 Sensor:
Environmental factors may influence the function or accuracy of the pulse oximeter,
such as ambient light, physical movement, diagnostic testing, low perfusion, electro-
magnetic interference, dysfunctional haemoglobin, presence of certain dyes and
inappropriate positioning of the pulse oximeter probe.
A functional test cannot be used to assess the Accuracy of a Pulse Oximeter Probe
or a Pulse Oximeter.
sensor accessory, the ventilator displays
2
How to Use the Vivo 45 LS
module
2
module.
2
69