6.3.3

High PEEP Alarm

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
114
Alarms
User manual
Description
High PEEP
Medium
A High PEEP alarm will be given when the measured
PEEP is 30% above the set value for more than 15 seconds
Blocked leakage port.
Too short expiratory time.
Changes in airway resistance and or compliance.
Malfunction of the exhalation valve.
Blocked exhalation valve.
PEEP has gone below the alarm limit (lower than 30%
above the set value).
The Vivo 45 LS will continue treatment according to the
current settings.
On
Off
Doc. 007301 K-1