Spo2 Disconnected; Nection Alarm); Spo 2 Signal Lost Alarm - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

6.4.8
SpO
Property
Alarm text
Alarm condition
Priority
Possible cause
Ventilator action
Reset
6.4.9
SpO
Property
Alarm text
Alarm condition
Priority
Possible cause
Ventilator action
Reset
Doc. 007301 K-1
Disconnected (SpO
2
Description

SPO2 Disconnected

An SpO
when an error signal or no signal from the SpO
been detected for 2 seconds. Check the SpO
High
The SpO
sequently no communication (possibly due to disconnection)
for 2 seconds.
Failure in the SpO
The ventilator will continue treatment with the same settings.
Confirmation by user or reconnected/changed.
Signal Lost Alarm
2
Description
SPO2 Signal Lost
SpO
High
Signal lost reported by SpO
removing the probe from finger, or sensor detached from
SpO
The ventilator will continue treatment with the same settings.
User presses OK or electronics cable is disconnected by the
user, or the sensor is reconnected to the finger.
Sensor Failure/Disconnection Alarm)
2
Sensor Failure/Disconnection alarm will be given
2
electronics cable has been disconnected and sub-
2
sensor.
2
signal lost.
2
electronics.
2
sensor has
2
sensor.
2
electronics (due to patient
2
User manual
Alarms
141

Advertisement

Table of Contents
loading

Table of Contents