5.5.10

Sigh %

Item
Definition
Setting min
Setting max
Setting
resolution
Default
value
NOTE
In pressure modes (during the sigh breath), the high pressure alarm will automatically
be set 10 cmH
5.5.11

Rise Time

Item
Definition
Setting min
Setting max
Setting
resolution
Doc. 007301 K-1
Description
Sigh % sets the increased % of the set pressure is delivered to the
patient.
125% of actual set pressure or volume .
200% of actual set pressure or volume . The maximum sigh is also
limited by the max allowed set volume.
25%
125%
O above set sigh pressure (max 70 cmH
2
Description
The Rise Time setting controls the speed of the pressure/volume
increase from start of inspiration to the set pressure or volume.
A low setting will give a faster increase and therefore a longer plateau
at the set value. A high setting will give a slower increase and therefore
a shorter plateau.
In the VCV-SIMV mode, two different Rise Time settings are made,
one for mandatory ventilator-controlled breaths and one for addi-
tional pressure support breaths triggered by the patient.
1 (PSV and PCV modes)
50% of the inspiration time (Min. 0.3 s) (VCV modes)
9 (PSV and PCV modes)
90% of the inspiration time (Min. 0.3 s) (VCV modes)
Off (VCV modes)
1 step(PSV and PCV modes)
10% of the inspiration time (VCV modes)
O).
2
How to Use the Vivo 45 LS
User manual
83