5.5.14

Exp. Trigger (Expiratory Trigger)

Item
Definition
Setting min
Setting max
Setting
resolution
5.5.15

Max Insp. Time (Maximum Inspiratory Time)

Item
Definition
Modes
Setting min
Setting max
Setting
resolution
Default
value
Doc. 007301 K-1
Description
The Expiratory Trigger setting defines the moment when the ventila-
tor will cycle from the inspiratory to the expiratory phase.
1 (10% decrease of peak flow)
9 (90% decrease of peak flow)
1 (Setting 1 is the most sensitive and 9 is the least sensitive)
Description
The Maximum Inspiratory Time setting defines a maximum length
for each inspiration. If the Maximum Inspiratory Time is set to Off,
the length of the inspiration and/or minimum inspiratory time is
dependent on the set Expiratory Trigger.
PSV, PSV(TgV)
0.3 s
5 s (Adult mode)
2 s (Pediatric mode)
Off
0.1 s
Off