High Inspco 2 Alarm - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

6.3.25
High InspCO
Property
Alarm text
Priority
Alarm condition
Possible cause
Setting range
Setting resolution
Ventilator action
This alarm requires a connected EtCO
134
Alarms
User manual
Alarm (High Inspired CO
2
Description
High InspCO2
High
A High Inspired CO
inspired CO
Alarm limit is set too low.
Excessive dead space between patient and exhalation valve/
leakage port.
Leakage port/valve occluded.
1 to 99 mmHg
Off
1 mmHg
The ventilator will continue treatment with the same settings.
)
2
alarm will be given when the measured
2
exceeds the alarm limit for 30 seconds.
2
sensor
2
Doc. 007301 K-1

Advertisement

Table of Contents
loading