High Mv E (High Expired Minute Volume Alarm) - Breas VIVO 65 User Manual

Hide thumbs Also See for VIVO 65:
Table of Contents

Advertisement

6.3.10
High MV
Item
Alarm text
Priority
Alarm condition
Possible cause
Setting range
Setting resolution
Ventilator action
Doc. 006150 M-1
(High Expired Minute Volume Alarm)
e
Description
High MVe
Medium
A High Expired Minute Volume alarm will be given when the
monitored expired minute volume exceeds the alarm limit for
15 seconds.
This alarm is only used if the Vivo 65's patient circuit type is set
to Dual limb or Leakage (single limb) circuit.
Mismatch between Breath Rate, Tidal Volume settings and
the alarm setting.
Increased Breath Rate.
Adult mode: 1.0 to 40 l/min
Paediatric mode: 1.0 to 20 l/min
Off
0.5 l/min
The ventilator will continue treatment with the same settings.
Alarms
99
User manual