6.3.4

Low PEEP Alarm

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Doc. 007301 K-1
Description
Low PEEP
Medium
A Low PEEP alarm will be given when the measured PEEP
is 30% below the set value for more than 60 seconds
Excessive leakage.
Malfunction of the exhalation valve.
PEEP has gone above the alarm limit (higher than 30%
below the set value).
The Vivo 45 LS will continue treatment according to the
current settings.
On
Off
Alarms
115
User manual