Low Vti Alarm (Low Inspired Tidal Volume)

Property
Alarm text
Priority
Alarm Condition
Possible cause
Setting range
Setting resolution
Reset criteria
Ventilator action
Doc. 007301 K-1
Description
Low Vti
High
A Low Inspired Tidal Volume alarm will be given when the
monitored Inspired Tidal Volume fails to reach the set limit
for the Low Inspired Tidal Volume alarm for 15 seconds.
Mismatch between Inspired Tidal Volume and Alarm setting.
Changes in airway resistance and or compliance.
Adult mode: 100 ml to 2000 ml
Pediatric mode: 20 ml to 500 ml
Off
10 below 600 ml, 100 above 600 ml
A full breath above set alarm limit
The ventilator will continue treatment with the same settings.