Breas Vivo 45 LS User Manual page 153

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

3 Connect the airway adapter to the CO
6.5.2.5 SpO
These tests applies if the SpO
1 Connect SpO
2 Set the low SpO
3 Set the high SpO
4 Start treatment and wait 30 s.
⇒ High SpO
5 Stop treatment.
6 Set the high SpO
7 Set the low SpO
8 Start treatment and wait 30 s.
⇒ Low SpO
9 Stop Treatment.
10 Set the low SpO
11 Set the low pulse rate alarm to off.
12 Set the high pulse rate alarm to 30 bpm.
13 Start treatment and wait 30 s.
⇒ High pulse rate alarm should be given.
14 Stop treatment.
15 Set the high pulse rate alarm to off.
16 Set the low pulse rate alarm to be 230 bpm.
17 Start treatment and wait 30 s.
⇒ Low pulse rate alarm should be given.
18 Stop Treatment.
19 Set the low pulse rate alarm to off.
6.5.2.6 Power Related Alarm Tests
1 If having the Click-in battery installed, disconnect it.
2 Check that the internal battery is fully charged and disconnect the Mains power
supply while the treatment is running.
⇒ The Lost mains power alarm shall be given.
Doc. 007301 K-1
Related Alarm Tests
2
sensor to device and to your finger.
2
alarm to 85%.
2
alarm to be 90%.
2
alarm should be given.
2
alarm to off.
2
alarm to be 100%.
2
alarm should be given.
2
alarm to 85%.
2
sensor again.
2
accessory is used,
2
Alarms
153