Frequencies Of Portable And Mobile Transmitters For Which The Recommended Separation Distance Is 30 Cm (12 Inches); Recommended Separation Distances Between External Power; Conductors And The Ventilator - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

8.3.4
Frequencies of portable and mobile transmitters for which the
recommended separation distance is 30 cm (12 inches)
Band (MHz)
380 — 390
430 — 470
704 —787
800 — 960
1,700 — 1,990
2,400 — 2,570
5,100 — 5,800
8.3.5

Recommended separation distances between external power

conductors and the ventilator

Rated maximum current in
conductor (A)
1
10
30
For conductors rated at a maximum current not listed above, the recommended separa-
tion distance d in meters (m) can be estimated using the equation d=I/2πH, where I is
the maximum current rating of the conductor in amperes (A) according to the transmit-
ter manufacturer; H is the ventilator immunity compliance level to electromagnetic fields
in the 50-60 Hz frequency span (30 A/m).
Doc. 007301 K-1
Service
TETRA 400
GMRS 460, FRS 460
LTE Band 13, 17
GSM 800/900, TETRA 800, iDEN 820,
CDMA 850, LTE Band 5
GSM 1800; CDMA 1900; GSM 1900;
DECT; LTE Band 1, 3, 4, 25; UMTS
Bluetooth, WLAN 802.11 b/g/n,
RFID 2450, LTE Band 7
WLAN 802.11 a/n
Separation distance (m)
50-60 Hz d= I/2πH= I/188
0.005
0.05
0.16
Immunity test level
(V/m)
27
28
9
28
28
28
9
Technical Specifications
User manual