Air Temp. Sensor Fail Alarm; Internal Error Alarm; Database Integrity Fail Alarm - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

6.4.24

Air Temp. Sensor Fail Alarm

Property
Alarm text
Alarm condition
Priority
Ventilator action
6.4.25

Internal Error Alarm

Property
Alarm text
Priority
Alarm Condition
Ventilator action
Reset action
6.4.26

Database Integrity Fail Alarm

Property
Alarm text
Priority
Alarm Condition
Ventilator action
Reset action
148
Alarms
User manual
Description
Air Temp Sensor Fail
The alarm is given in case of swivel boot temperature sensor
communication failure or sensor reporting temperatures out
of range (below -30°C (-22°F) or above 60°C (140°F).
Medium
The ventilator will continue treatment with the same settings.
Description
Internal Error
High
An internal Error alarm will be given when the ventilator has
an internal error, followed by an error code for the specific
failure. All Internal Error alarm error codes are defined and
explained in the ventilator Service Manual.
The ventilator will continue treatment with the same settings.
Power off and restart the ventilator.
Description
Database Integrity Failed
High
This alarm is given when the database integrity check fails.
The ventilator will continue treatment with the same settings.
Rebuild the database and restart the ventilator.
Doc. 007301 K-1