High Pulse Rate Alarm - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

6.3.26

High Pulse Rate Alarm

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
This alarm requires a connected SpO
Doc. 007301 K-1
Description
High Pulse Rate
Medium
A High Pulse Rate alarm will be given when the measured
pulse rate exceeds the alarm limit for 15 seconds.
Insufficient ventilatory support.
Too low flow of bleed-in oxygen.
The PEEP value is set too high.
Bad positioning of the finger probe.
The pulse rate goes back below the alarm limit.
The Vivo 45 LS will continue treatment according to the
current settings.
30 to 230 bpm (beats per minute)
Off
5 bpm (beats per minute)
sensor.
2
Alarms
135
User manual