6.3.18

Obstruction Alarm

Property
Alarm text
Priority
Alarm condition
Ventilator action
Reset Criteria
Setting Range
6.3.19

High FiO

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Setting range
Setting resolution
Ventilator action
Doc. 007301 K-1
Description
Obstruction
High
An Obstruction alarm will be given if the inspiratory breath-
ing tube becomes blocked and remains blocked for 2 con-
secutive breaths.
With each breath cycle, upon detection of an obstruction
the ventilator will reduce the airway pressure to the set
PEEP.
Treatment will resume with the start of the next breath cycle.
When the monitored compliance and resistance become
normal after a breath.
High
Low
Off

Alarm

2
Description
High FiO2
Medium
A High FiO
exceeds the alarm limit for 30 seconds.
Increased oxygen inflow.
Decreased minute ventilation.
FiO
goes below the alarm limit
2
21% to 100%
Off
1%
The ventilator will continue treatment with the same settings.
alarm will be given when the measured FiO
2
2
Alarms