Audible alarm level
Range/Performance: 1 to 5, where 1 is the lowest volume setting and 5 is the highest
volume setting.
Resolution: 1
8.2.3
Monitored Values Specifications
This section describes the ranges and tolerances for monitored values on the Vivo 45 LS.
All stated tolerances includes measurement uncertainty. The accuracies have been tested
with all allowed configurations. Stated tolerances only disclose the maximum tolerance.
P
peak
Range/Performance: 4 to 99 cmH
Resolution: ±0.5 cmH
PEEP
Range/Performance: 0 to 99 cmH
Resolution: ±0.5 cmH
P
mean
Range/Performance: 0 to 99 cmH
Resolution: ±0.5 cmH
Leakage
Range/Performance: 0 to 99.9 l/min (BTPS*).
Resolution: ±10%
MV
i
Range/Performance: 0 to 99.9 l (BTPS*).
Resolution: ±10% or (±15 ml × bpm), whichever is greatest
MV
e
Range/Performance: 0 to 99.9 l (BTPS*).
Resolution: ±15% or (±15 ml × bpm), whichever is greatest
168
Technical Specifications
User manual
O.
2
O or ±10%, whichever is greatest
2
O.
2
O or ±10%, whichever is greatest
2
O.
2
O or ±10%, whichever is greatest
2
Doc. 007301 K-1