Main Unit; Air Pathway Disinfection; Patient Circuit - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

7.1.1

Main Unit

1 Switch off the Vivo 45 LS and disconnect the power supply.
2 Remove the patient circuit.
3 Disconnect all electric cables.
4 Clean the outside of the Vivo 45 LS using a lint-free cloth with a mild soap solution,
and/or ethanol 70% for surface disinfection.
5 Reconnect the patient circuit. Make sure all parts are dry before the ventilator is put
into operation.
7.1.2

Air Pathway Disinfection

The table below lists the parts that might get contaminated by exhaled gases or bodily
fluids during normal use or single fault condition.
Condition
With bacteria filter
Without bacteria filter
In case of contamination, the internal air pathways of the Vivo 45 LS may be disinfected
up to 10 times by a maximum 60 minute long validated ozone gas process.
Low resistance bacteria filter, if used, should be replaced every 24 hours.
7.1.3

Patient Circuit

The patient circuit should be cleaned and replaced in accordance with the manufac-
turer's instructions and care provider's instructions, where applicable. For safety
information, read 2.4 Usage of Patient Circuit, page 26.
Doc. 007301 K-1
Parts
Patient circuit
EtCO
airway adapter (if used)
2
Bacteria filter
Patient circuit
EtCO
airway adapter (if used)
2
FiO
sensor (if used)
2
Patient air outlet/Pneumatic unit
Air bypass unit
Blower/Inlet silencer
Air inlet with filters
Cleaning and Maintenance
User manual
157