6.3.11
High MV
Item
Alarm text
Priority
Alarm condition
Possible cause
Setting range
Setting resolution
Ventilator action
122
Alarms
User manual
(High Expired Minute Volume Alarm)
e
Description
High MVe
Medium
A High Expired Minute Volume alarm will be given when
the monitored expired minute volume exceeds the alarm
limit for 15 seconds.
•
Mismatch between Breath Rate, Tidal Volume settings and
the alarm setting.
•
Increased Breath Rate.
•
Adult mode: 1.0 to 40 l/min
•
Pediatric mode: 1.0 to 20 l/min
•
Off
0.5 l/min
The ventilator will continue treatment with the same settings.
Doc. 007301 K-1