High Mv E (High Expired Minute Volume Alarm) - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

6.3.11
High MV
Item
Alarm text
Priority
Alarm condition
Possible cause
Setting range
Setting resolution
Ventilator action
122
Alarms
User manual
(High Expired Minute Volume Alarm)
e
Description
High MVe
Medium
A High Expired Minute Volume alarm will be given when
the monitored expired minute volume exceeds the alarm
limit for 15 seconds.
Mismatch between Breath Rate, Tidal Volume settings and
the alarm setting.
Increased Breath Rate.
Adult mode: 1.0 to 40 l/min
Pediatric mode: 1.0 to 20 l/min
Off
0.5 l/min
The ventilator will continue treatment with the same settings.
Doc. 007301 K-1