5.5.2

Insp. Pressure

Item
Definition
Setting min
Setting max
Setting
resolution
Default
value
5.5.3

PEEP

Item
Definition
Setting min
Setting max
Setting
resolution
Default
value
Doc. 007301 K-1
Description
The Insp. Pressure setting is used to define the airway pressure during
the inspiratory phase.
Minimum/maximum working IPAP is limited/achieved by a software
control of blower speed vs. measured pressure.
4 cmH
O
2
50 cmH
O
2
(limited to 30 cmH
O above PEEP in Pediatric mode)
2
0.5 below 10 cmH
O, 1.0 above 10 cmH
2
15 cmH
O
2
Description
The PEEP setting is used to define the airway pressure at the end of
the expiratory phase.
2 cmH
O
2
Off (for patient circuits with active exhalation valve)
20 cmH
O
2
For pressure ventilation modes: The max setting is also limited by
Insp. Pressure -2 cmH
2
0.5 below 10 cmH
O, 1.0 above 10 cmH
2
5 cmH
O (Adult mode)
2
2 cmH
O (Pediatric mode)
2
O
2
O and Min Pressure -2 cmH
O
2
How to Use the Vivo 45 LS
O.
2
79
User manual