7.3.5
High Minute Volume Alarm
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
7.3.6
Low Minute Volume Alarm
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Doc. 007231 F-3
Description
High MV
Medium
A High Minute Volume alarm will be given when the minute
volume is above the set alarm limit for 15 seconds.
•
Disconnection of patient circuit.
•
Mismatch between pressure setting and alarm setting.
•
Leakage from the mask or other components of the patient
circuit.
•
Increased breath rate.
The minute volume goes below the alarm limit.
The Vivo 3 will continue treatment according to the current
settings.
Description
Low MV
High
A Low Minute Volume alarm will be given when the minute
volume is below the set alarm limit for 15 1 seconds.
•
Disconnection of patient circuit.
•
Mismatch between pressure setting and alarm setting.
•
Leakage from the mask or other components of the patient
circuit.
•
Decreased breath rate.
The minute volume goes above the alarm limit.
The Vivo 3 will continue treatment according to the current
settings.
Alarms 59
User Manual