WARNING!
Risk of Electric Shock
High voltage contact may cause cardiac arrhythmia.
Repairs and modifications must be carried out by authorized technicians only and in
accordance with instructions from Breas Medical
CAUTION!
Do not attempt to autoclave or sterilize the Vivo 45 LS.
2.8

Usage of Oxygen

When using the Vivo 45 LS with oxygen, always follow the oxygen provider's instruc-
tions and use only medical grade oxygen complying with local regulations.
WARNING!
As this medical device uses an alternative small-bore connector design different
from those specified in the ISO80369 series, there is a possibility that a misconnec-
tion can occur between this medical device and a medical device using a different
alternative small-bore connector, which can result in a hazardous situation causing
harm to the patient. Special measures need to be taken by the user to mitigate these
reasonable foreseeable risks.
WARNING!
Risk of fire
The presence of oxygen can speed up combustion of inflammable materials.
Risk of Fire
When oxygen is used with the Vivo 45 LS, the oxygen flow must be turned off when
the Vivo 45 LS is not operating. Oxygen delivered into the patient tubing may accu-
mulate within the machine enclosure. Oxygen accumulated in the machine enclosure
increases the risk of fire.
Doc. 007301 K-1
The Vivo 45 LS must not be opened, repaired or modified by unauthorized per-
sonnel. If subjected to unauthorized operations, Breas Medical is no longer
responsible for the device's performance and safety and all warranties will become
invalid.
The Vivo 45 LS must not be modified or interconnected to unapproved
equipment.
Safety Information
User manual
31