4 Start treatment on both the ventilator and the CPAP device.
5 Wait approximately 15 seconds before the High PEEP alarm shall be given.
6 Stop treatment. Test completed.
6.5.3.2 Low Pressure and Disconnection Alarms
1 Start treatment and disconnect the patient circuit.
2 Wait 15 seconds.
3 The Low Pressure Alarm and/or the Disconnection Alarm will be given.
4 Stop treatment. Test completed.
6.5.3.3 Disconnection Alarm Test
1 Set the disconnection alarm to On.
2
⇒ The disconnection alarm shall be given.
3 Set the disconnection alarm to Off.
6.5.3.4 Obstruction Alarm
1 Start treatment; block the patient circuit completely to simulate an obstruction.
2 Wait approximately 10 seconds.
3 The Obstruction Alarm will be given.
4 Stop treatment. Test completed.
Doc. 007301 K-1
Disconnect the patient circuit.
Alarms
155