6.4.3
Low Patient Air Temp. (Low Patient Air Temperature Alarm)
Property
Alarm text
Alarm condition
Priority
Possible cause
Ventilator action
6.4.4

Low Last Power Source Alarm

Property
Alarm text
Priority
Alarm condition
Ventilator action
6.4.5

Crit. Low Last Power Source Alarm

Property
Alarm text
Alarm condition
Priority
Ventilator action
Reset
Doc. 007301 K-1
Description

Low Patient Air Temp.

A Low Patient Air Temperature alarm will be given when
the patient air temperature is below the preset limit -30°C
(-22°F).
High
Too low ambient temperature
The ventilator will continue treatment with the same settings.
Description
Low Last Power Source
Medium
This alarm will be given when the last battery source (inter-
nal battery) has 15 minutes of operating time left with cur-
rent settings.
The Vivo 45 LS will continue treatment according to the
current settings.
Description
Crit. Low Last Power Source
A Crit. Low Last Power Source alarm will be given when the
last battery source (internal battery or click-in battery) has
5 minutes of operating time left with current settings.
High
The ventilator will continue treatment with the same settings.
Connection of "higher" power source.
Alarms
139
User manual