Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Doc. 007301 K-1
Description
Disconnection
High
A Disconnection alarm will be given when the measured
flow exceeds the expected leakage flow at the set Pressure
for 15 seconds.
•
Too high leakage in the patient circuit.
•
The patient has removed the mask.
•
Circuit disconnection.
•
Pilot pressure tube disconnection
The leakage is back within limits.
The Vivo 45 LS will continue treatment according to the
current settings
•
On
•
Off
Alarms
127
User manual