Breas Vivo 45 LS User Manual page 127

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Doc. 007301 K-1
Description
Disconnection
High
A Disconnection alarm will be given when the measured
flow exceeds the expected leakage flow at the set Pressure
for 15 seconds.
Too high leakage in the patient circuit.
The patient has removed the mask.
Circuit disconnection.
Pilot pressure tube disconnection
The leakage is back within limits.
The Vivo 45 LS will continue treatment according to the
current settings
On
Off
Alarms
127
User manual