Functions And Parameters In The Vivo 45 Ls; Parameters' Availability In Different Modes - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

In MPV breath mode it is possible to set Breath Rate to zero so that breaths are only ini-
tiated when the patient triggers them using the mouthpiece. The PEEP setting is not
available but always set to Off so that no air blows from the mouthpiece when no breath
is being delivered.
As the patient is not always connected to the ventilator several of the surveillance func-
tions are not working in the same way as in other modes:
Disconnection alarm is not available in MPV mode.
Low Pressure alarm is only active during breaths.
Apnea alarm is possible to set to longer times and is an important mean of surveil-
lance to make sure that the patient gets ventilation support regularly.
When switching to MPV mode all alarms, except from High/Low Pressure Alarms, are
automatically switched off to avoid false alarm triggering (when changing between pro-
files the alarm settings does not change to Off, but stays as defined in the profiles).
WARNING!
The alarm levels must be assessed and adjusted considering the patient's condition
and treatment settings.
MPV breath mode shall be used with the mouthpiece circuit including the mouth-
piece interface only.
The MPV circuit including the mouthpiece interface shall be used in MPV breath
mode only and is not for use in any other breath mode.
MPV breath mode shall not be used with ventilator dependent patients.
5.4.4.13 CPAP – Continuous Positive Airway Pressure
In CPAP mode the Vivo 45 LS is applying a continuous positive pressure to the airways.
The flow will automatically be adjusted to maintain the set CPAP level.
5.5

Functions and Parameters in the Vivo 45 LS

All the parameters used for controlling the breathing by the Vivo 45 LS are listed below.
Depending on the setup for Home mode adjustment, parameters may not be available
or have limited ranges when the Vivo 45 LS is in Home mode.
5.5.1

Parameters' Availability in Different Modes

This table offers an overview of the available parameters in each mode.
76
How to Use the Vivo 45 LS
User manual
Doc. 007301 K-1

Advertisement

Table of Contents
loading

Table of Contents