Target Volume
Range/Performance:
Off, 300 to 2000 ml (Adult), Off, 50 to 500 ml (Pediatric).
Tolerance: ±12 ml or ±10%, whichever is greatest.
Resolution:
10 ml below 500 ml
50 ml above 500 ml
Max Pressure
Range/Performance: Min IPAP to 50 cmH
Resolution: 0.5 below 10 cmH
Min IPAP
Range/Performance: 4 cmH
Resolution: 0.5 below 10 cmH
Tidal Volume
Range/Performance:
300 to 2000 ml (Adult), 50 to 500 ml (Pediatric).
Tolerance: ±12 ml or ±10%, whichever is greatest.
Resolution: 10 ml below 500 ml, 50 ml above 500 ml
Flow Pattern , VCV-SIMV
Range/Performance: Square, Decelerating
CPAP
Range/Performance:
4 to 20 cmH
Tolerance: ±0.5 cmH
Resolution: 0.5 below 10 cmH
Doc. 007301 K-1
O, 1.0 above 10 cmH
2
O to Max Pressure.
2
O, 1.0 above 10 cmH
2
O.
2
O or ±5%, whichever is greatest.
2
O, 1.0 above 10 cmH
2
O.
2
O
2
O
2
O
2
Technical Specifications