6.4.27

Cooling Fan Error Alarm

Property
Alarm text
Alarm Condition
Priority
Ventilator action
Reset
6.4.28

Clock Failure Alarm

Property
Alarm text
Priority
Alarm condition
Ventilator action
Reset action
6.4.29

Internal Temp High Alarm

Property
Alarm text
Priority
Alarm condition
Ventilator action
Reset
Doc. 007301 K-1
Description
Cooling Fan Error
The Cooling Fan Error alarm shall be given when the cooling
fan runs too slow.
High
The ventilator will continue treatment with the same settings.
When cooling fan speed is above 275 rpm.
Description
Clock Failure
High
The alarm shall be given when the real time clock value is
invalid.
The ventilator will continue treatment with the same settings.
Restart the ventilator.
Description
Internal Temp High
High
The Internal High Temp alarm shall be given when the venti-
lator internal temperature is high.
The internal temp high alarm is triggered when PTU/Sensor
board temperature is higher than 65°C (149°F), or main
board temperature is higher than 65°C (149°F), or motor
temperature is higher than 85°C (185°F).
The ventilator will continue treatment with the same settings.
When the triggering conditions are resolved.
Alarms
149
User manual