5.5.19

Flow Pattern

ITEM
Definition
Settings
Default
value
5.5.20

Max Pressure

Item
Definition
Setting min
Setting max
Setting
resolution
Tolerance
Default
value
88
How to Use the Vivo 45 LS
User manual
DESCRIPTION
Flow pattern sets the characteristics of the air flow in VCV modes.
Square (constant flow during the inspiratory phase)
Decelerating (flow decreases linearly, may prevent air hunger)
Square
Description
The Max Pressure setting is only used when Target Volume is
activated.
Max Pressure defines the upper pressure limit up to where the ventila-
tor can increase the pressure to reach the set Target Volume. If Target
Volume is not reached at Max Pressure, the ventilator will continue to
ventilate at this Max Pressure setting.
Current Min pressure setting
50 cmH
O
2
0.5 below 10 cmH
O, 1.0 above 10 cmH
2
±0.5 cmH
O or ±5 %, whichever is the greatest.
2
15 cmH
O
2
O
2