6 Start the treatment.
6.5.2.1 High and Low Pressure Alarm Tests
1 Set the high pressure alarm to 10 cmH
⇒ The high pressure alarm shall be given.
2 Set the high pressure alarm to 55 cmH
3 Set the low pressure alarm to 20 cmH
⇒ The low pressure alarm shall be given.
4 Set the low pressure alarm to 1.0 cmH
6.5.2.2 Expiratory Tidal Volume Alarm (Vt
This alarm test applies if having a patient circuit with intentional leakage.
1 Set up the ventilator as described in Alarm Test Preparation, page 151.
2 Set the high Vt
⇒ The high Vt
3 Set the high Vt
4 Set the low Vt
The low Vt
6.5.2.3 Inspiratory Tidal Volume Alarm (Vt
This alarm test applies if having a patient circuit with exhalation valve or a patient circuit
with mouthpiece.
1 Set up the ventilator as described in Alarm Test Preparation, page 151.
2 Set the high Vt
⇒ The high Vt
3 Set the high Vt
4 Set the low Vt
The low Vt
6.5.2.4 EtCO
2
This alarm test applies if the EtCO
1 Connect the EtCO
2 Disconnect the airway adapter from the CO
⇒ The check CO
152
Alarms
User manual
alarm to 150 ml.
e
alarm shall be given.
e
alarm to Off.
e
alarm to 400 ml.
e
alarm shall be given.
e
alarm to 150 ml.
i
alarm shall be given.
i
alarm to Off.
i
alarm to 400 ml.
i
alarm shall be given.
i
Related Alarm Test
sensor with an attached airway adapter to the Vivo 45 LS.
2
adapter alarm shall be given.
2
O.
2
O.
2
O.
2
O.
2
) Tests
e
)Tests
I
accessory is used.
2
sensor.
2
Doc. 007301 K-1