6.3.8
Low MV
Property
Alarm text
Priority
Alarm condition
Possible cause
Setting range
Setting resolution
Ventilator action
Doc. 007301 K-1
(Low Inspired Minute Volume Alarm)
i
Description
Low MVi
High
A Low Inspired Minute Volume alarm will be given when
the monitored minute volume does not reach the alarm limit
for 15 seconds.
•
Mismatch between Breath Rate and Inspired Tidal Volume
settings and the alarm setting.
•
Changes in airway resistance and or compliance.
•
Decreased Breath Rate.
•
1.0 l/min to 30 l/min (Adult mode)
•
0.1 l/min to 10 l/min (Pediatric mode)
•
Off
0.1 l up to 1.0 l, 0.5 l above 1.0 l.
The ventilator will continue treatment with the same settings.
Alarms
119
User manual