Maintenance
Ion Source Interface Maintenance
86
LTQ Series Hardware Manual
To check the Convectron gauge pressure
1. Open the Tune Plus window (see
2. Choose Setup > Vacuum to open the Vacuum dialog box

Figure 58. Vacuum dialog box

3. Check the reading for the Convectron Gauge Pressure. If the reading is considerably
below 1 Torr, clean the ion transfer tube.
To remove the ion transfer tube
1. Turn off the flow of liquid from the LC (or other sample introduction device) to the API
source as follows:
a. Open the Tune Plus window (see
b. Choose Setup > Inlet Direct Control to open the Inlet Direct Control dialog box
(Figure
59).

Figure 59. Inlet Direct Control dialog box for an Accela 1250 Pump

c. Click the Pump Off or Stop Pump button.
page
34).
(Figure
page
34).
Click this button to
turn off the pump.
58).