System Setup

Connecting the Dionex ICS-4000 Power Cord

Thermo Scientific
This chapter provides instructions for setting up the system hardware and Chromeleon
software. These instructions apply to systems configured with a conductivity detector or with
an electrochemical detector. After completing the setup steps, go on to the plumbing
instructions for your detection type.
Contents
Connecting the Dionex ICS-4000 Power Cord
Plugging in the EGC and CR-TC
Installing Cartridges in the Dionex IC Cube
Installing the Detector
Connecting an Autosampler
Connecting to the Computer
Setting Up the Chromatography Software
Connecting the Drain, Waste, and Vent Lines
Degassing the Water
Setting Up the Reservoir
Priming the Pump
Checking the Conductivity of the Water
To connect the power cord
1. Verify that the power switch on the rear panel of the Dionex ICS-4000 is turned off. The
power switch may be turned on accidentally when the module is unpacked.
2. Connect the power cord (IEC 320 C13) (ordered separately) from the main power
receptacle on the rear panel to a grounded power source. The IC system power supply is
auto-sensing, so no adjustment is required to select the line voltage.
Dionex ICS-4000 Installation Instructions
2
5