Coil Temperature Adjustment

IMPORTANT
This procedure applies only to models P7000HD, P7X06H, P7X08H,
P7X15, and P7220.
Do this procedure only when the original controller board or shuttle frame
assembly have been replaced or you are instructed to do so in a maintenance
or troubleshooting procedure. Do not do this procedure if the original
controller board or shuttle frame assembly were removed and reinstalled as
part of other maintenance tasks.
IMPORTANT
The shuttle frame assembly must be at room temperature to do this
procedure. Power off the printer and let it cool for at least one hour
before doing this procedure.
1. Plug the AC power cord into the printer and the power source.
2. Power on the printer.
3. The printer must be offline to do this procedure. If the printer is online,
4. On the control panel, press the
5. Press
6. Press ENTER. The display tells you to "PLEASE WAIT" while the coil
7. When the display again reads "Factory / Set Coil Temp" press
8. Press the ON LINE key to exit the Factory menu.
9. Coil temperature is set. Return the printer to normal operation (page 231).
press the ON LINE key. "OFFLINE / QUICK SETUP" will display.
"ENTER SWITCH UNLOCKED" briefly displays. (If "LOCKED" displays,
=
>
simply press
and
again. This is the default key combination. The lock/
unlock key combination can be programmed by the user. If
not unlock ENTER, get the new key combination from the user.)
=
>
<
;
+
+
+
the same time.) "Factory / Set Coil Temp" appears on the display.
temperature is set by an automatic calibration sequence in printer
software.
lock the ENTER key.
Adjustments and Tests
=
>
+
keys to unlock the ENTER key.