6 Maintenance Procedures

Fitting the source enclosure to the instrument

Required materials
Chemical-resistant, powder-free gloves
To fit the source enclosure to the instrument:
Warning:
source enclosure to the source if a corona pin is fitted. The pin's tip is
sharp.
1.
Using both hands, fit the source enclosure to the two supporting studs
on the source adaptor housing.
Notice:
NanoLockSpray source enclosure, you must slide the sprayer
platform out of the source enclosure before you close the enclosure
(see
2.
Close the source enclosure.
3.
Connect the probe adjuster and options cables to the instrument's
connectors.
The cables and connectors are color coded; the blue-sleeved cable
Tip:
connects to the blue connector and the yellow-sleeved cable to the yellow
connector.
4.
Slide closed the instrument's source interface door.
102
April 2, 2015, 715004496 Rev. B
To avoid personal contamination with biologically
Warning:
hazardous, toxic, or corrosive materials, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when working
with the LC system connections, ESI probe, and source
components.
To avoid puncture wounds, take great care while fitting the
To avoid damaging the sample inlet, when fitting a
page
72).