Advertisement

4.
Disconnect the probe adjuster and options cables from the instrument's
connectors.
Warning:
working with the source enclosure open if a corona pin is fitted.
The pin's tip is sharp.
To avoid damaging the sample inlet, when removing a
Notice:
NanoLockSpray source enclosure, you must slide the sprayer
platform out of the source enclosure before you open the enclosure
(see
page
5.
Pull the source enclosure release (located at the bottom, right-hand side)
outward, and swing open the enclosure.
6.
Using both hands, grasp the source enclosure, and lift it vertically off the
two supporting studs on the source adaptor housing.
Source enclosure
7.
Store the cables neatly by plugging them into the cable-storage positions
on the rear of the source enclosure.
Removing and refitting the source enclosure
To avoid puncture wounds, take great care while
72).
Cable storage positions
Supporting stud
TP03164
April 2, 2015, 715004496 Rev. B