5 Installing and removing the ionKey source

Removing the ionKey source

You can remove the ionKey source, and replace it with a conventional
interface.
Alternative:
M-Class system mounted on an M-Class cart fitted with an ionKey source
holder, you can secure the source enclosure to the holder. Doing so keeps the
enclosure close to the Xevo G2-XS, for when it is next needed; assists with
managing the ionKey source's fluid lines; and helps prevent contamination of
the fluid lines.
See the ACQUITY M-Class documentation for further information about
installing the ionKey source holder on the M-Class cart, and securing the
source enclosure to the holder.
Recommendation:
cable-management bracket and a replacement infill panel. You can leave the
bracket and infill panel in place when you replace the ionKey source with a
conventional interface. If, however, you choose to remove the bracket and
replace the original infill panel, reverse the installation instructions
step
9) in
"Installing the ionKey source" on page
"Installing the ionKey source" on page
See also:
Required materials
Chemical-resistant, powder-free gloves
¼-inch wrench
Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves while performing this procedure. The source
components can be contaminated.
Warning:
performed on its source before beginning this procedure.
88
April 2, 2015, 715004496 Rev. B
If you are using the ionKey source with an ACQUITY UPLC
The ionKey source is installed along with a
To avoid personal contamination with biohazards or
To avoid electric shock, prepare the instrument for work
78.
78.
(step 5
to