Waters Xevo G2-XS QTof Overview And Maintenance Manual page 188

Hide thumbs Also See for Xevo G2-XS QTof:
Table of Contents

Advertisement

6 Maintenance Procedures
To install the new capillary:
1.
Use the sharp knife or PEEK tubing cutter to cut an approximately
60-cm (24-inches) length of red PEEK tubing.
Requirement:
2.
Insert one end of the red, PEEK tubing in the probe inlet connector, and
screw the connector, finger-tight, into the PEEK union.
Rationale:
capillary.
3.
Fit the UNF coupling to the new capillary.
188
April 2, 2015, 715004496 Rev. B
To avoid personal contamination with biologically
Warning:
hazardous, toxic, or corrosive materials, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when working
with the probe and source components.
Cut the tubing perpendicular to its horizontal axis.
Doing so ensures a minimum dead volume when fitting the
Probe inlet connector
PEEK tubing
TP02671