6 Maintenance Procedures
4.
Remove the end cover and gasket.
Nebulizer adjuster knob
5.
Unscrew and remove the nebulizer adjuster knob.
6.
Remove the PEEK union/UNF coupling assembly and capillary from the
probe.
The PEEK union used with the IonSABRE II probe is notched on
Tip:
one of its flats, a feature that distinguishes it from the PEEK union used
with the ESI probe (see
Notch
186
April 2, 2015, 715004496 Rev. B
page
PEEK union/UNF coupling assembly
Capillary
Locknut
Gasket
End-cover
175).