Waters Xevo G2-XS QTof Overview And Maintenance Manual page 113

Hide thumbs Also See for Xevo G2-XS QTof:
Table of Contents

Advertisement

To empty the nitrogen exhaust-trap bottle:
1.
To stop the LC Flow, follow the action below for your software:
Software
MassLynx
UNIFI
2.
Pull the source enclosure release (located at the bottom, right-hand side)
outward, and swing open the enclosure.
3.
Unscrew and remove the nitrogen exhaust-trap bottle from the cap and
associated fittings.
4.
Dispose of the waste liquid in accordance with local environmental
regulations.
5.
Fit and tighten the nitrogen exhaust-trap bottle into the cap.
6.
Secure the nitrogen exhaust-trap bottle in the upright position.
7.
Close the source enclosure.
An automatic pressure test is performed.
Tip:
Action
In the Instrument Console, click Stop Flow
On the System Console tool bar, click Stop Flow
To avoid personal contamination with
Warning:
biologically hazardous, toxic, or corrosive materials,
and to avoid spreading contamination to
uncontaminated surfaces, wear clean,
chemical-resistant, powder-free gloves when handling
waste liquid that accumulates in the nitrogen exhaust
trap.
To avoid contaminating uncontaminated
Warning:
surfaces with biologically hazardous, toxic, or corrosive
materials, dispose of waste liquid according to local
environmental regulations.
Emptying the nitrogen exhaust-trap bottle
April 2, 2015, 715004496 Rev. B
.
.
113