Waters Xevo G2-XS QTof Overview And Maintenance Manual page 276

Hide thumbs Also See for Xevo G2-XS QTof:
Table of Contents

Advertisement

B External Connections
Required materials
Chemical-resistant, powder-free gloves
7-mm nut driver
Sharp knife
The following items are included in the Xevo G2-XS installation kit:
NW25 center rings
NW25 clamps
12.7-mm clear PVC exhaust tubing
PVC hose clamps
Vacuum hose
PTFE drip tray
To connect the roughing pump:
1.
Make sure there is adequate ventilation around the pump, so that the
ambient temperature around the pump does not exceed 40°C. Allow a
minimum 150 mm clearance for pump cooling fans.
276
April 2, 2015, 715004496 Rev. B
To avoid personal contamination with
Warning:
biologically hazardous, toxic, or corrosive materials, and
to avoid spreading contamination to uncontaminated
surfaces, wear clean, chemical-resistant, powder-free
gloves when working with the pump and its
connections.