8.
Unscrew the finger-tight nut on the rear of the reference sprayer
assembly.
9.
Remove the liner tubing and capillary from the reference sprayer
assembly.
10. Remove the capillary from the liner tubing.
11. Dispose of the capillary in accordance with local environmental
regulations.
Replacing the LockSpray reference probe capillary
To avoid contaminating uncontaminated
Warning:
surfaces with biologically hazardous, toxic, or corrosive
materials, dispose of the capillary according to local
environmental regulations.
April 2, 2015, 715004496 Rev. B
193